Call Us Now
+86 15260605085
Environment test chamber can be used:
1. as a stand-alone test for environmental effects on test specimens
2. as preparation of test specimens for further physical tests or chemical tests.
3. as environmental conditions for conducting testing of specimens
Item No.:
E001N-408LPayment:
TT, L/C, Wester UnionProduct origin:
ChinaShipping port:
China PortLead Time:
10-35 daysBrand:
UTSTESTERCertificates:
CE, ISO 17025Standards:
IEC 60068-2-1;Overview
Constant temperature and humidity test chamber, also known as Temperature & Humidity Environmental Test Chambers, Humidity test chamber, programmable hot and humid alternating test chamber, constant temperature machine or constant temperature and humidity chamber, used to test the performance of materials in a variety of environments and test a variety of materials heat resistance, cold resistance, dry resistance, moisture resistance performance. Suitable for electronics, electrical appliances, cell phones, communications, instruments, vehicles, plastic products, metal, food, chemical, building materials, medical, aerospace and other products for testing quality
UTS temperature humidity chamber available to be programmed for different temperature and humidity.
Standards
GB/T2423.2-89 High-temperature test method Bb
|
Features
1. With USB download function, with RS232 interface, curved software, no need for recorder
2. With vacuum glass observation window, more effective direct observation function
3. Refrigerant servo valve flow calculation control. Energy saving 30%, with a dormant function of the resident flow time, energy saving 50% or more
4. Anti-electromagnetic interference controller with SGS certification; with more than 32 numbers of soft program matching, with multiple development and software upgrade function
5. ISO-9001 quality certification (Taiwan & Dongguan & Suzhou)
6. Defrost cycle (800 defrost cycles once, each defrost guaranteed within 2 hours, integrated test to achieve 25% energy saving)
Optional Models
Model |
E001-80L |
E001-100L |
E001-150L |
E001-225L |
E001-408L |
E001-800L |
E001-1000L |
Inner box volume |
80L |
100L |
150L |
225L |
408L |
800L |
1000L |
Test chamber size (D*W*H) (cm) |
40×40×50 |
40×50×50 |
50×50×60 |
50×60×75 |
60×80×85 |
80×100×100 |
100×100×100 |
Weight(kg) |
180 |
200 |
250 |
300 |
550 |
960 |
1100 |
Key Specifications
Temperature |
|
The temperature range |
-20 ℃ ~ + 150 ℃ / -40℃ ~ + 150 ℃ / -70℃ ~ + 150 ℃ |
Temperature deviation |
≦ ± 2.0 ℃ |
Temperature fluctuations |
± 0.5 ℃ |
Temperature evenness |
≦ 2.0 ℃ |
cooling rate |
Heating rate: - 20 ℃ to + 100 ℃ about 35 min (At no-load state, environmental temperature + 25 ℃) Cooling rate: from + 20 ℃ to - 20 ℃ about 45 min (At no-load state, environmental temperature + 25 ℃) |
Humidity |
|
Humidity range |
20%R.H~98%R.H |
Humidity deviation |
±3.0%RH(>75%RH) ±5.0%RH(≤75%RH) |
Humidity uniformity |
±3.0%RH (No load) |
Humidity fluctuation |
±2.0%RH |
The temperature and humidity performance test is based on the relevant regulations of IEC 60068-3 standard.The sensor is placed at the outlet of the unit. |
Material to be Tested